Moisture Analysis: PET Resin
| Sample Type | Required Quantity | Equipment Used |
| Resin Flake Preform Bottle |
1 grams | Karl Fisher Coulometric Titrator and Vaporizer |
Purpose
This procedure measures the amount of moisture within a sample using
the coulometeric Karl Fischer technique.
Description
A sample is deposited into the analyzer and heated to release all of
the moisture within the sample. The moisture then reacts with the iodine
and generates a current which allows for an accurate measurement to be
made. Once the analyzer moisture reading falls to within the background
reading, the analyzer calculates the ppm moisture by the calculated
weight of water (reacted with the iodine) divided by the input sample
weight.
← Go back to Plastic Materials Testing

PTI delivers innovation, development and comprehensive technical services to progressive companies engaged in plastic packaging.

Contact
Phone: +41 24 423 9530
Fax: +41 24 423 9531
E-Mail