Water Bottles

Moisture Analysis: PET Resin

 

Sample Type Required Quantity Equipment Used
Resin
Flake
Preform
Bottle
1 grams Karl Fisher Coulometric Titrator and Vaporizer

 

Purpose
This procedure measures the amount of moisture within a sample using the coulometeric Karl Fischer technique.

Description
A sample is deposited into the analyzer and heated to release all of the moisture within the sample. The moisture then reacts with the iodine and generates a current which allows for an accurate measurement to be made. Once the analyzer moisture reading falls to within the background reading, the analyzer calculates the ppm moisture by the calculated weight of water (reacted with the iodine) divided by the input sample weight.

 

← Go back to Plastic Materials Testing

Our Mission


PTI delivers innovation, development and comprehensive technical services to progressive companies engaged in plastic packaging.


 

Contact

Phone: +41 24 423 9530
Fax: +41 24 423 9531
E-Mail